Ashcroft 1495 - Low Pressure Receiver Gauge Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Instrumentos de medida Ashcroft 1495 - Low Pressure Receiver Gauge. Ashcroft 1495 - Low Pressure Receiver Gauge User Manual [fr] Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir

Resumo do Conteúdo

Página 1 - MAINTENANCE

PRESSURE GAUGEINSTALLATION,OPERATION ANDMAINTENANCEI&M008-10098-5/02 (250-1353H) 1M AMR 5P1/08

Página 2

10TYPICAL CALIBRATION CHARTINDICATED VALUE (PSI)True Value – Increasing – Increasing – Decreasing – Decreasing –PSI Without RAP With RAP Without RAP W

Página 3 - CONTENTS

116.1 Calibration – Rotary Movement Gauges– Inspect gauge for accuracy. Many timesgauges are simply “off zero” and a simplepointer adjustment using th

Página 4

12BOURDON TUBESEGMENTTIPLINKBACKPLATESOCKETROTARY MOVEMENTPINIONHAIRSPRINGFig. 2TYPE 1082 EXTERNAL ZERO ADJUST FEATURE*ARINGINSTRUCTIONS FOR USE:LOOSE

Página 5

13TYPE 1082 TEST GAUGE CALIBRATION PROCEDUREFig. 4

Página 6 - 2.0 TEMPERATURE

14d. Remove pointer and reassemble dial anddial screws (finger tight).e. Lightly press pointer onto pinion.f. Check accuracy at full scale. If error e

Página 7 - 4.0 OPERATION

1525Vs1009 DURALIFE®PRESSURE GAUGE CALIBRATION

Página 8 - PROCEDURES/DEFINITIONS

167.0 DIAPHRAGM SEALS7.1 General – A diaphragm seal (isolator) is adevice which is attached to the inlet connec-tion of a pressure instrument to isola

Página 9

17tions must pass before they reach the bourdontube. The dampener reduces the magnitude ofthe pressure pulse thus extending the life ofthe Bourdon tub

Página 10 - TYPICAL CALIBRATION CHART

189.0 RESOURCES9.1 Training Videos9.1.1 Test gauge calibration9.1.2 1009 Duralife®gauge calibration9.1.3 Duragauge®gauge calibration9.1.4 Diaphragm se

Página 13

Visit our web site www.ashcroft.comAshcroft Inc.,250 East Main StreetStratford, CT 06614-5145U.S.A.Tel: 203-378-8281Fax: 203-385-0408 (Domestic)Fax: 2

Página 14

3PageCONTENTS1.0 Selection and Application1.1 Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 15 - PRESSURE GAUGE CALIBRATION

4CONTENTS7.0 Diaphragm Seals7.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 16 - 8.0 DAMPENING DEVICES

1.0 SELECTION & APPLICATIONUsers should become familiar with ASMEB40.1 (Gauges – Pressure Indicating DialType – Elastic Element) before specifying

Página 17

62.0 TEMPERATURE2.1 Ambient Temperature – To ensure longlife and accuracy, pressure gauges shouldpreferably be used at an ambient tempera-ture between

Página 18 - 9.0 RESOURCES

7also be drained from the case and the front ringloosened before autoclaving.3.0 INSTALLATION3.1 Location – Whenever possible, gaugesshould be located

Página 19

4.5 Other considerations – These include (a)bent or unattached pointers due to extremepressure pulsation; (b) broken windows whichshould be replaced t

Página 20 - 1M AMR 5P1/08

9Calibration chart – After recording all of thereadings it is necessary to calculate the errorsassociated with each test point using the fol-lowing fo